Full list of precautions (continued)
Always place the cables into a
washer/disinfector tray before
loading the tray into the washer/
disinfector to avoid damage.
Do not use pressurised or medical
air at greater than 3 bar (43.5 psi) to
avoid damage to the electrosurgery
instrument cables.
Do not coil the cables too tightly
when packing into a peel pouch as
this could result in damage to the
cable. The coil diameter should be at
least 16 mm.
Full list of information points
Follow all applicable national and
local laws and guidelines or hospital
protocols for cleaning/disinfection
and sterilisation.
Always follow the manufacturer’s
recommended concentrations and
application times for the detergent
and disinfectant.
Begin point-of-use preparation within
30 minutes of the surgical procedure
ending.
Do not use abrasive metal brushes
as these may damage the Versius
Electrosurgery Instrument Cables.
Transport and store cables in line
with local guidelines.
Start initial manual cleaning within
4 hours of nishing point-of-use
preparation.