Sterilisation
Warnings
Check that items placed in the
sterilisation tray are not touching or
shadowing each other as this could
result in ineective sterilisation.
Ensure that electrosurgery
instrument cables have been
thoroughly dried before sterilisation
as any moisture remaining could
result in ineective sterilisation.
Always place packaged cables into a
sterilisation tray before placing the
tray into the steam steriliser to avoid
damage.
Only use validated parameter
settings. Parameter settings that
have not been validated could
result in damage to, or ineective
sterilisation of, Versius Electrosurgery
Instrument Cables.
The sterilisation process should
conform with ISO 17665-1.
Allow electrosurgery instrument
cables to cool down gradually to
room temperature before handling.
Sudden changes in temperature
could damage electrosurgery
instrument cables.
Precautions
Appendix A lists the maximum
number of reprocessing cycles for
Versius Electrosurgery Instrument
Cables. Do not reprocess Versius
Electrosurgery Instrument Cables
beyond their maximum number of
reprocessing cycles as this may result
in damage to the cables.
Take care when handling and
transporting Versius Electrosurgery
Instrument Cables to avoid damage.
63