Main cleaning:
Full manual cleaning
Warnings
Only use validated cleaning solutions.
Cleaning solutions that have not been
validated (e.g. bleach or hydrogen
peroxide) may result in damage to, or
ineective cleaning of, cables.
Take care when handling
contaminated uids or chemicals, as
skin contact, inhalation or ingestion
may cause harm. Personal Protective
Equipment (PPE) should be worn,
according to local guidelines.
Do not use strong acidic (pH < 7)
solutions, cleaners or rinse aids
as these may damage Versius
Electrosurgery Instrument Cables.
Do not use strong alkaline (pH > 11)
solutions, cleaners or rinse aids
as these may damage Versius
Electrosurgery Instrument Cables.
During full manual cleaning, always
rinse the cables under running
critical water for at least 1 minute 30
seconds after exposing the cables
to pH 7–11 enzymatic detergent, as
some detergent or contaminants
could remain on the cables.
During full manual cleaning, always
brush the cables under cold running
water using a soft-bristled brush until
there is no visible soil.
During full manual cleaning, ensure
that the cable is fully covered in
pH 7–11 enzymatic detergent when
soaking the cable.