Precautions
Do not use an ultrasonic bath when
reprocessing the electrosurgery
instrument cables. Using an
ultrasonic bath will cause damage to
the cables.
Appendix A lists the maximum
number of reprocessing cycles for
Versius Electrosurgery Instrument
Cables. Do not reprocess Versius
Electrosurgery Instrument Cables
beyond their maximum number of
reprocessing cycles as this may result
in damage to the cables.
Information points
Follow all applicable national and
local laws and guidelines or hospital
protocols for cleaning/disinfection
and sterilisation.
Always follow the manufacturer’s
recommended concentrations and
application times for the detergent
and disinfectant.
Do not use abrasive metal brushes
as these may damage the Versius
Electrosurgery Instrument Cables.
38
Main cleaning: Full manual cleaning