Main cleaning:
Automated cleaning and
thermal disinfection
Warnings
Only use validated cleaning solutions.
Cleaning solutions that have not been
validated (e.g. bleach or hydrogen
peroxide) may result in damage to, or
ineective cleaning of, cables.
Take care when handling
contaminated uids or chemicals, as
skin contact, inhalation or ingestion
may cause harm. Personal Protective
Equipment (PPE) should be worn,
according to local guidelines.
Do not use strong acidic (pH < 7)
solutions, cleaners or rinse aids
as these may damage Versius
Electrosurgery Instrument Cables.
Do not use strong alkaline (pH > 11)
solutions, cleaners or rinse aids
as these may damage Versius
Electrosurgery Instrument Cables.
Always use ISO 15883 compliant
washer/disinfectors.
Check that items placed into the
washer/disinfector tray are not
touching or shadowing each other.
This could result in ineective
cleaning/disinfection of the items.
Only use validated parameter
settings. Parameter settings that
have not been validated could result
in damage to, or ineective cleaning
of, Versius Electrosurgery Instrument
Cables.
45