EasyManua.ls Logo

CMR versius - Page 46

Default Icon
80 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Main cleaning: Automated cleaning and thermal disinfection
Precautions
Do not use an ultrasonic bath when
reprocessing the electrosurgery
instrument cables. Using an
ultrasonic bath will cause damage to
the cables.
Always place the cables into a
washer/disinfector tray before
loading the tray into the washer/
disinfector to avoid damage.
Appendix A lists the maximum
number of reprocessing cycles for
Versius Electrosurgery Instrument
Cables. Do not reprocess Versius
Electrosurgery Instrument Cables
beyond their maximum number of
reprocessing cycles as this may result
in damage to the cables.
Information points
Follow all applicable national and
local laws and guidelines or hospital
protocols for cleaning/disinfection
and sterilisation.
Always follow the manufacturer’s
recommended concentrations and
application times for the detergent
and disinfectant.
46
Main cleaning: Automated cleaning and thermal disinfection

Other manuals for CMR versius