Loading...
Point-of-use preparation
Information points
Follow all applicable national and
local laws and guidelines or hospital
protocols for cleaning/disinfection
and sterilisation.
Always follow the manufacturer’s
recommended concentrations and
application times for the detergent
and disinfectant.
Begin point-of-use preparation within
30 minutes of the surgical procedure
ending.
Start initial manual cleaning within
4 hours of nishing point-of-use
preparation.