Step 3 Soak cable
A 1. Fill a sink or container with enough fresh pH 711 enzymatic detergent to
completely immerse the cable.
2. Place the cable in the sink or container.
COVER WITH
B Leave the cable to soak for a minimum of 10 minutes (unless the detergent
manufacturer’s instructions recommend a longer soaking period).
SOAK 10 MINUTES
MINIMUM
C When the soaking period is complete, remove the cable from the sink or
container.
D Drain the sink or container of the detergent.
Continue to Step 4 on page 34