Sterilisation
Step 1 Load sterilisation tray into steam steriliser
Load the sterilisation tray into the
steam steriliser according to the steam
steriliser manufacturer’s instructions.
LOAD INTO
Step 2 Run steriliser
Set parameters and run sterilisation cycle (according to ISO 17665-1):
Prevacuum steam sterilisation parameters
Parameter Value
Temperature 134–137 °C (273–279 °F)
Exposure time 3–18 minutes (minimum holding time)
Dry time 20 minutes (minimum)
OR
Parameter Value
Temperature 132 °C (270 °F)
Exposure time 4 minutes (minimum holding time)
Dry time 20 minutes (minimum)
SET AND RUN
Continue to Step 3 on page 67
The sterilisation process should
conform with ISO 17665-1.
66
Sterilisation