a single beep will be heard and you may
remove your finger from the POWER key. The
display will change to OSP.
2. Rotate the center knob until the display reads
PL. Press the “Heat” key under the HEAT dis-
play to accept, and the display will change to
250 (250°C limit), or UL (unlimited). Rotate the
center knob to change the display to the method
of choice and press the “Heat” key again.
3. The unit will return to the previous menu. Select
another feature to change, or press the POWER
key again to return to the off mode.
Probe Response - Pr
This feature is available for units with heating functionality
and is only applicable when using a probe. The purpose
of this feature is to select the minimum time required for a
temperature change to be detected before signaling a
Probe Out of Solution error (E03). A choice between 3
and 20 minutes in 1 minute increments is possible. The
default is 3 minutes, but if a large load is placed on the
top, the time may need to be extended to avoid nuisance
E03 errors.
To modify/view the Probe Response follow the instruc-
tions given below:
1. Plug the unit into the appropriate power, but do
not turn it on (displays should be blank, unless
Hot Top Warning System is indicating a hot sur-
face), enter the Service Menu by pressing and
holding the POWER key. After about 3 seconds
a single beep will be heard and you may
remove your finger from the POWER key. The
display will change to OSP.
2. Rotate the center knob until the display reads
Pr. Press the “Heat” key under the HEAT dis-
play to accept, and the display will change to
the current time chosen. Rotate the center knob
to change the display and press the “Heat” key
again.
SERVICE AND CALIBRATION