19
TROUBLESHOOTING
egasseMdeya l ps i Dtce t eDo tdedne t n IesuaCnoi tuloS
70Eneewt ebecne r e f f i deg r aL
dnae l puocomr eh t l an r e t n I
.elpuoco mrehtPT O
elpuocomrehtP TOrolanretnI
.detcennocton
elpuocomrehtP TOrolanretnI
.tiuc rictr ohs
elpuocomrehtP TOrolanretnI
sd r awkcabde t cennoc
.)yt iralopdesrever(
dnano i t cennocr epo r pe r usnE
de t ce f f af oy t i r a l op
.elpuoco mreht
de t ce f f af ot r ohsevomeR
.elpuoco mreht
dnano i t cennocr epo r pe r usnE
de t ce f f af oy t i r a l op
.elpuoco mreht
11Eylrepor pt onrewopCA
.detceted
no i t ce t edr ewopCAn ie r u l i aF
.tiuc ric) sso r corez (
.draoBlor tnoCecalpeR
21Egni rr i tsnirotordekcoL
.lortnoc
.noi t idnocrotordekcoL
.rotomnier ul iaF
/relpuocotporotomnierul iaF
.tiuc riccairt
laci tporotomnierul iaF
.leehwredocne/redocne
.rotordekcoleer F
.rotomecalpeR
.draoBlor tnoCecalpeR
laci tporotomecalpe R
.redocne
12E.yro m e mhsal fatad detpur roCatadgni rud erul iaf m uskcehC
.yrevocerhsal f
roec i vreSremotsuCt catnoC
.wo l ebe t on t oo fees
suoun i t noch t iwya l ps i Dkna l B
.peeb
.yromemmargorpdetpur ro Ctinugni ruderul iafmuskceh C
.noi tazi lai t ini
.draoBlor tnoCecalpeR
.draoBlor tnoCm argorpeR
Errors E11 and E12 are stirring errors. Error Handler will lock out stirring functions if stirring error is detected.
To avoid boil over, the unit will cease heating (the user can restart heating if desired). Pressing the POWER
button or unplugging the unit will clear Error E12. Unplugging the unit will clear Error E11.
Error E21 is a data flash memory error and can be corrected by following the steps below:
Press any key to clear the error, this re-initializes data flash memory.
At the ‘SEL menu, press the “Heat” or “Stir” key.
Using the encoder knob, select your unit’s model type (if your unit model number starts with SP,
select ‘ SP’, etc) and press either the “Heat” or “Stir” key.
Using the encoder knob, select your unit’s top size (either 7 or 10 inch) and press either the “Heat”
or “Stir” key.
Press the POWER key to exit the special functions menu.
Note that it may be necessary to recalibrate the unit and/or re-enter custom settings as the unit has
been reset to factory default.
The unit can now be started normally using the POWER key.