Appendix B Quick Commands
Restore Default Settings
1. Press the Settings button on the Screen Selection tool bar.
2. Press the Restore Defaults button on the settings screen.
3. A prompt will appear warning the user that selecting OK will return all settings
to a default state.
Force Relays to shutoff when a process is started
1. Press the Settings button on the Screen Selection tool bar.
2. In the Relay Control section, check Start Shutoff for the desired relay you wish
to shutoff on process start.
Set when relays activate during deposition process
1. Press the Settings button on the Screen Selection tool bar.
2. In the Relay Control section, select the round radio button associated with the
step during the deposition process when the relay should activate.
Manual: The relays will not activate during the deposition process
automatically, but can still be controlled by the Relay # button.
Start: At the start of each layer/material in the process the relay will activate
and will deactivate at the end of each material.
Predeposit: The relay will activate during the predeposit phase of the
deposition process for each layer/material. The relay will then shutoff at the
end of the deposition process.
Auto Deposition: The relay will activate during the dwell phase, just before
the PID activation. This allows the dwell time to occur between shutters
release and PID activation.
Mirage: Eon™ uses Relay 2 to activate Mirage™ when the Tempe™
heater is engaged. Mirage™ MUST be selected before the heater can be
used, allowing Eon™ to control Relay 2 automatically during the deposition
process.
Setting when Eon™ automatically zeros sensor thickness
1. Press the Settings button on the Screen Selection tool bar.
2. In the Sensor Zeroing menu, checkmark each setting associated with the
sensor that is to have its thickness automatically zeroed.
Before Log: Eon™ zeros thickness when Log button is pressed.
Start Pressed: Eon™ zeros thickness when Start” button is pressed to start
a new deposition.
Before Layer: Eon™ zeros thickness each time a new layer/material
engages during the process.
Before PID: Eon™ zeros thickness each time a new layer/material
activates the PID.