14
English
m
i-SCAN
• This function adapts the electrostimulation
session to the physiology of each user.
Just before starting the work session,
m
i-
SCAN probes the chosen muscle group
and automatically adjusts the stimulator
parameters to the excitability of this area of the
body. It is a truly personalised measurement.
N.B.: The
m
i-SCAN function operates as
soon as a stimulation cable tted with
the
m
i-SENSOR system (sold separately) is
connected to the stimulator.
• This function is implemented at the start
of the programme by a short sequence
during which measurements are taken (a
horizontal bar scans the gurine at the left
of the screen).
• Throughout the entire measurement test,
you must remain completely still and be
perfectly relaxed. The
m
i-SENSOR system is
very sensitive: the smallest contraction or
the slightest movement could disturb the
measurement test.
• Note that certain people might feel an
unpleasant tingling sensation during the
test.
• Once the test is complete, the symbol
will appear, indicating that the
programme can be started.
m
i-TENS
• The
m
i-TENS function can be used
to considerably reduce the number
of unwanted muscular contractions,
enhancing both user comfort and
treatment ecacy.
• The
m
i-TENS function can only be
accessed for the Acute tendinitis and
Ankle twist programmes.
• For these programmes, the
m
i-TENS
function allows you to control the
stimulation energies while maintaining
them at an eective level and
considerably limiting the onset of
muscular contractions.
• Short measurement tests (lasting 2 to
3 seconds) are carried out regularly
throughout the entire programme.
• There is a test phase after each increase
of the stimulation energies. To ensure the
correct performance of these tests, you
must remain perfectly still during this
time.
• Depending on the results of the
measurement tests recorded by the device,
the level of the stimulation energies may
be slightly decreased automatically.
• It is important always to adopt the most
comfortable possible stimulation position.
Also, you should try to remain still and not
contract the muscles of the stimulated
area.
m
i-RANGE
• This function indicates the ideal energy
adjustment range for programmes whose
eciency requires vigorous muscular
twitches.
• The
m
i-RANGE function is therefore only
available in programmes that use low
stimulation frequencies (less than 10 Hz).
• For programmes allowing the use of the
m
i-RANGE function, the stimulator checks
whether you are in the ideal energy range.
If you are below the range, the stimulator
prompts you to increase the energy by
displaying
signs.
• Once the stimulator has detected your
ideal adjustment area, a square bracket
will appear to the right of the bar graph
of the channel to which the
m
i-SENSOR
system is connected. The gures in this
bracket indicate the energy range you
must work in order to achieve optimal
results.
• If you set the energy level below this
range, the stimulator will prompt you to
increase it by ashing a series of
signs on the screen.

Table of Contents

Other manuals for Compex Runner

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Compex Runner and is the answer not in the manual?

Compex Runner Specifications

General IconGeneral
BrandCompex
ModelRunner
CategorySwitch
LanguageEnglish

Related product manuals