Copyright © 2022, Condor Instruments Ltda. 17
Professor Tulio Street Ascarelli, 290 - Vila Madalena, São Paulo, SP - Cep:05449-020
Manufacturer's guide and statement —
Immunity and electromagnetic emissions
Medical electrical equipment requires special precautions regarding the
Electromagnetic Compatibility (CEM) and also needs to be installed and put into operation according to the
information regarding the EMC contained in this document.
Guide and Manufacturer's Declaration - Electromagnetic Emissions
The device is intended for use in the electromagnetic environment specified below. It is advisable that the purchaser or user of the device ensures
that it is used in such an environment.
Electromagnetic environment - guide
RF emissions
(Radio frequency) CISPR11
The device only uses radio frequency energy for its inner workings. For this reason, RF
emissions are very low and are unlikely to cause interference to electronic equipment in
their vicinity.
RF emissions
(Radio frequency) CISPR11
The device is suitable for use in all locations, including homes and locations connected
directly to the low-voltage public network that provides power to
Harmonic emissions
IEC 61000-3-2
Fluctuations in voltage/
Emissions with jitter
IEC 61000-3-3
Guide and Manufacturer's Declaration - Electromagnetic Immunity
The device is intended for use in the electromagnetic environment specified below. It is advisable that the purchaser or user of the device ensures
that it is used in such an environment.
Electromagnetic environment - guide
Electrostatic Discharge (DES)
IEC
61000-4-2
± 8 kV Per contact
± 2 kV, ± 4 kV, ± 8
kV, ± 15 kV Via or
± 8 kV Per contact
± 2 kV, ± 4 kV, ± 8
kV, ± 15 kV Via or
The floors should be made of wood, concrete or ceramic. If the
floors are covered by synthetic material, the relative humidity
should be at least 30%.
Fast/salvac electrical
transients
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for
input/output lines
The device is charged via computer and is not directly subject to
the power supply network
± 1 kV line(s) the
line(s)
± 2 kV line(s) solo
The device is charged via computer and is not directly subject to
the power supply network
Voltage outages, short
interruptions and voltage
variations in the input lines of
the power supply
IEC 61000-4-11
<5% Ut (>95% drop
in Ut) during 0.5
cycle
40% Ut (60% drop in
Ut) for 5 cycles
70% Ut (30% drop in
Ut) for 25 cycles
<5% Ut (>95% drop
in Ut) during
5
The device is charged via computer and is not directly subject to
the power supply network
Magnetic field generated by
the frequency of the electrical
network (50/60 Hz)
IEC 61000-4-8
Magnetic fields in the frequency of the power supply network
should have levels characteristic of a typical location in a typical
hospital or commercial environment.
NOTE 1: Ut is the voltage of alternating current before applying the t level