Manual EV2000/EV3000 series Electrophoresis Power Supply, rev. 03/2021
Page 7
Method programming
Method programming permits to store up to 9 different methods, each with 9 steps. Each step is able to
recall a next one, providing a flexible multiple step function for special techniques.
The method mode also permits to program a linear voltage gradient for any step provided the limiting current
or power is not attained.
Parameters can be changed temporarily without interrupting the run.
Verify if the instrument is switched off.
Connect the electrophoresis tank to the instrument.
Switch the instrument on.
If the timer is still running after switching on, the power supply will automatically proceed with a
procedure for the remaining time. A temporary menu is shown: Choose YES within 15 sec. to stop
the procedure. Choose NO or wait 15 sec. to continue the procedure for the remaining timer value.
Select METHOD and press :
Select the desired method and press :
Choose YES to modify the name of the
procedure, NO to proceed with step 8.
To modify the name, select each
character from left to right, max. 12
characters.
Select the space character to end a
name with less than 12 characters
and return to point 6.
Selectable characters are:
a to z, A to Z’, ‘0 to 9, _’, ‘,’ , ‘-‘, ‘.’
When previously multiple steps have
been programmed, select the desired
step and press :
Decide whether or not a linear voltage
gradient has to be performed
and press :
METHOD 1|1
MODIFY NAME?
Method1
NO
METHOD 1|1
ADJUST NAME
Method1
METHOD 1|1
STEP
1
METHOD 1|1
GRADIENT
NO
PROCEDURE?
MANUAL
METHOD
SETUP
CHOOSE
METHOD
1