RECOMMENDED MAINTENANCE SCHEDULE
BEFORE EVERY USE
Confirm that cables are tight and secure.
Visually check lenses for dust.
AFTER EVERY USE
Place lens caps on cameras.
Back up any critical tests to safe media.
EVERY 1 MONTH
Perform a full dust check per the instruction by viewing a blank white wall.
Dust front and back of lens and sensor glass with clean compressed air.
Check calibration grids for any heavy damage or smudges; clean with soap and water as necessary.
Back up any important tests from hard drive to safe or offsite media.
EVERY 6 MONTHS
Inspect all cables for crimps/damage.
Disconnect both ends of camera cables and check for damaged connectors.
Open PC and carefully dust motherboard and fans.
EVERY 12 MONTHS
Consider renewing support contract this will extend hardware warranty for as long as the contract is