product.
Make sure that the surface of the product is absolutely dry before plugging the power cable back into the
socket.
After interventions on and repairs of the instrument and before re-use, have the service personnel
perform safety checks on the instrument.
5. Accessories and Combinations with Other Instrument
Use of unauthorized accessories or unauthorized modifications of the instrument could lead to injury.
Only use accessories that have been approved for combination with the product by the manufacturer.
Only use accessories that are equipped with standardized interfaces.
Do not make any modifications to the instrument unless these have been approved by the manufacturer
of the product.
6. Electromagnetic Fields
Electromagnetic fields might interfere with the functions of implanted systems (such as pacemakers). Medical
electrical instruments are subject to special precautions regarding electromagnetic compatibility and must be
installed and operated in accordance with the tables of electromagnetic compatibility. About electromagnetic
compatibility refer to 12
High-frequency communications instruments may interfere with medical electrical instruments.
Ask patients if they have a cardiac pacemaker or other system implanted before you start the treatment.
Comply with the tables of electromagnetic compatibility during installation and commissioning.
If the instrument needs to be used in the immediate vicinity of another instrument, monitor the instrument
or system for malfunctions.
7. Contra-angle
Only use the original contra-angle.
Never press the contra-angle push button when handpiece is running. It will cause the file to fall off.
Never remove the contra-angle during operation.
Only use undamaged root canal instruments, refer to 7.
Never place your fingers on the moving parts of the instrument while it is running.
Before use, check the contra-angle for any damage or loose part.
8. Root canal instruments
Never use continuous rotary instruments in reciprocating mode.
Never use reciprocating instruments in rotary mode.
Refer to the file manufacturer’s instructions to adjust the speed and torque.
Description of the Product
Endo Motor products are mainly used in dental root canal preparation is used for each model pulpit and pulp
necrosis and various root tooth root canal treatment of important instrument.
1. Intended Use
The Endo Motors device is a endodontic treatment motorized handpiece with root canal measurement
capability. It can be used to enlarge the canals while monitoring the position of the file tip inside the canal. The
instrument must only be used in hospital environments, clinics or dental offices, by qualified practitioners.
2. Contraindications
In cases where a patient has been fitted with an implanted heart pacemaker (or other electrical