sensor type | Self-heated thermistor bridge with integral temperature compensation |
---|---|
operating range | Atmospheric to 0 microns |
accuracy | +/- 10% of Reading |
operating temperature range | 0°C to 50°C (32°F to 122°F) |
---|---|
storage temperature | -40 to 85°C (-40 to 185°F) |
operating & storage humidity | 0-95%, non-condensing |
operating pressure range | Atmospheric pressure to 0 microns |
---|---|
maximum working pressure | 600 PSIG |
burst pressure | 3000 PSIG |
power source | One 9 volt alkaline battery |
---|---|
battery life | 20 hours of continuous use |
battery indicator | Battery symbol with 3 segment Power Level Bar Graph |
weight | 153 g (5.5 oz) excluding the battery |
---|---|
dimensions | 68 mm diameter by 40 mm deep by 124 mm overall height |
mechanical connection | Standard 1/4” female SAE refrigerant hose type with core depressor |
Details the VG200 as a state-of-the-art digital vacuum gauge, highlighting its ruggedness, reliability, ease of use, and advanced features.
Outlines various ways to connect the VG200 to service manifold gauge sets, refrigerant system ports, or vacuum pumps.
Explains how to turn on the unit, select measurement units, and initiate the evacuation process.
Describes how to use the VG200 to test vacuum pump performance before evacuating a system.
Details the self-heated thermistor sensor and its integral temperature compensation.
Specifies the operating pressure, burst pressure, and operating ranges for the unit.
Lists accuracy, operating temperature, and humidity specifications.
Covers power source, battery life, weight, dimensions, and mechanical connection details.