User Manual for Handheld Pulse Oximeter
21
F. Data Update
8 beats moving average for both SpO
2
and pulse rate readings
G. Data Storage
Recording SpO
2
and pulse rate data every 1/2/4/8 second(s),
up to 384–hour records can be stored.
H. Low Perfusion Performance
The accuracy of SpO
2
and PR measurement still meet the
precision described above when the modulation amplitude
is as low as 0.5%.
I. Resistance to interference of surrounding light:
The difference between the SpO
2
value measured in the
condition of indoor natural light and that of darkroom is
less than ±1%.
J. Dimensions: 145 mm (L) × 74 mm (W) × 29 mm (H)
Net Weight: 210g (including batteries)
K. Classification
Type of protection against electric shock:
Internally powered equipment
Degree of protection:
Type BF applied parts.
Degree of protection against harmful ingress of liquids:
Ordinary equipment without protection against ingress of
water.
Electro-Magnetic Compatibility: Group I, Class B