User Manual for Handheld Pulse Oximeter
24
8 Repair and Maintenance
8.1 Maintenance
The expecte
d
service life(not a warranty) of this device is 5
years. In orde
to ensure its long service life,
p
lease
p
ay
attention to the maintenance;
y
Please change the
b
atteries when the low-voltage indicato
r
appears.
y
Please take out the
b
atteries if the oximete
r
will not
b
e
used for a long time.
y
The recommende
d
storage environment of the device:
Ambient temperature: -20ºC ~60ºC
Relative humidity 10%~95%
Atmospheric pressure: 50kPa~107.4kPa
y
The oximete
r
is calibrate
d
in the factory
b
efore sale, there
is no nee
d
to calibrate it during its life cycle.
However, if it is necessary to verify its accuracy routinely,
the use
r
can do the verification
b
y means of SpO2
simulator, o
r
it can
b
e done
b
y the local thir
d
p
arty test
house.
8.2 Cleaning and Disinfecting Instruction
y
Surface-clean senso
r
with a soft cloth
b
y wetting with a
solution such as 75% isopropyl alcohol, i f low-level
disinfection is required, use a 1:10
b
leach solution.
y
Then surface-clean
b
y a dampene
d
cloth an
d
letitai
r
dry
or wipe it with a cloth.
High-pressure disinfection cannot be used on the device.
Do not immerse the device in liquid.