Gasman Operation
13
2.5 Logging
Gasman incorporates both data and event logging which can be
accessed using the RS232 communications link available with the Single
Way Charger Interface (part No.C01940), using Crowcon Portables PC
software. See section VIII.
Data is recorded every minute (this rate can be adjusted using the PC
software). The log can record 900 hours of data at 1-minute intervals.
Gasman also records the time and date for a number of operating and
diagnostic events including:
• Switch on and switch off;
• Level 1, Level 2 and TWA, alarm on, alarm off and the peak level
during the alarm;
• Zero, calibration and gas test with success or failure;
• Flammable sensor saver on and off;
• The battery condition is logged every switch on and switch off
while the instrument is operating, and certain configuration
changes are also logged;
• The Event log can record over 4800 events.