Pellistor saving Triple Plus+ and Triple Plus+ IR
Additionally, a pellistor will switch itself off if it reaches a reading of 105% LEL
or more – this action will happen irrespective of whether a % volume sensor is
also fitted. If a pellistor switches itself off then it will stay off until it is manually
switched on. This happens by pressing the unmarked button. The pellistor will
warm up for 5 seconds before the pellistor channel will read.
Pellistor saving is indicated on the display by the gas level being blank for the
flammable block.
3.5 Configuring The Instrument With Portables
PC
The following configuration settings may ONLY be set via Portables PC. The
function of these parameters is explained here. See the Portables PC documen-
tation for details of how these parameters are accessed and modified. Note that
Portables PC also allows the parameters configurable via the front panel to be
configured via a PC.
CONFIDENCE BLIP The green LED will always flash once every three seconds.
This option determines if an audible blip is also made.
LOG PERIOD Allows setting of the data logging period, anywhere from 1 sec-
ond to 24 hours. See section on data logging for description.
S.T.E.L. TIME The short term exposure limit (S.T.E.L.) averaging period, set
between 5 and 15 minutes. 15 minutes is the UK standard.
SYSTEM TIME AND DATE Instrument time and date.
CALIBRATION DUE Sets the date at which the Calibration due warning mes-
sage will appear at instrument start up.
USER AND SITE STRINGS The actual text associated with the 10 user and site
names may be set via Portables PC. The information, once set, appears in the
log menu.
VOLUME RANGE Set the volume of gas used by flammable volume channels
to turn off flammable (LEL) channels (for Autoranging units only).
PASSWORD Allows the menu password to be changed from the default of
1 2 3 4.
CHANNEL NAME AND UNITS This is the text used for the gas readings display.
It may be set via Portables PC.
FLAMMABLE CORRECTION DATA The name associated with a correction fac-