Specications 28
XP2i Operation Manual

PRESSURE RANGES, DISPLAY SCALES, & RESOLUTION
psi bar kPa Overpressure psi
in
H
2
0 inHg mmHg mmH
2
0 kg/cm
2
bar mbar kPa MPa
15PSI 1BAR 100KPA 6.5 x 0.001 0.01 0.001 0.01 1 0.0001 0.0001 0.1 0.01 0.00001
30PSI 2BAR 200KPA 3.0 x 0.001 0.01 0.001 0.1 1 0.0001 0.0001 0.1 0.01 0.00001
100PSI 7BAR 700KPA 2.0 x 0.01 0.1 0.01 0.1
1
0.0001 0.0001 0.1 0.01 0.00001
300PSI 20BAR 2KKPA 2.0 x 0.01 0.1 0.01 1 0.001 0.001 1 0.1 0.0001
500PSI 30BAR 3KKPA 2.0 x 0.01
1
0.1
1
0.001 0.001
1
0.1 0.0001
1KPSI 70BAR 7KKPA 2.0 x 0.1 0.1 0.001 0.001 0.1 0.0001
2KPSI 140BAR 14KKPA 2.0 x 0.1 0.1 0.01 0.01 1 0.001
3KPSI 200BAR 20KKPA 1.5 x 0.1 0.1 0.01 0.01 1 0.001
5KPSI 300BAR 30KKPA 1.5 x 0.1
1
0.01 0.01 1 0.001
10KPSI 700BAR 70KKPA 1.5 x 1 0.01 0.01 1 0.001
15KPSI 1000BAR 100KKPA 1.3 x 1 0.01 0.01 1 0.001
Unneeded pressure units may be disabled via the RS-232 connector using CongXP software.
kPa models can display pressure in kPa, MPa, and bar (or mbar) only. psi and bar models can display all available units.
XP2i will indicate pressure up to 10% above Range Pressure. Above 110%, the XP2i display will ash, indicating that the applied pressure exceeds the cali-
brated pressure range. If the calibrated pressure range is exceeded, the pressure displayed may not be accurate.
Absolute pressure XP2i's are designated by a “B” in front of “XP2i” in the part number.
An Absolute pressure version is available only on XP2is and XP2i-DDs with ranges of 2000 psi and higher.

PART NUMBERING SYSTEM
P/N Model Dual Display?
Accuracy
Adapter:
Type
Connection:
Location
Panel Mount? Data-logging?
XP2I
No .....(omit) 0.1% of reading ...(omit) NPT .......(omit) Bottom ..(omit) No .....(omit) No ..... (omit)
Yes .......-DD 0.02% of FS ..........-S2 G 1/4 B ..... -BSP Back ........-RP Yes ........ -F4 Yes ........ -DL
0.05% of FS ..........-S5 M20x1.5 ...-M20
"omit" for -F4
panel mount
Includes
-RP option
SAMPLE PART NUMBERS
300PSIXP2I ............300 psi standard gauge
200BARXP2I-DD .......200 bar dual-line display gauge
2KPSIBXP2I ............2000 psi Absolute gauge
700KPAXP2I-S2 .......700 kPa gauge with 0.02% of FS accuracy
-RP option Rear Pressure Fitting -F4 option Panel Mount Flange
CPF Adapter Fitting is not included.