Safety & Certifications
HAZARDOUS LOCATIONS
Every XP2i pressure gauge includes the following Intrinsic Safety approvals:
II 1G Ex ia IIC T4/T3 Ga
FTZU 12 ATEX 0048 X
Ex ia IIC
T4/T3 Ga
IECEx FTZU 12.0009 X
!
WARNINGS: The following warnings apply to the XP2i:
•
Use of Duracell MN1500 batteries is not approved for CSA certication.
•
Do not use the RS-232 connector in hazardous locations.
•
Replace batteries in non-hazardous locations and with approved types only.
•
Do not mix battery types or manufacturers.
•
T4, T3C, or T3 Temperature Class is based on selection of approved battery. See Approved Batteries on Page 27.
•
Special conditions for safe use:
• The equipment shall not be installed in a location where external conditions are conducive to the buildup of electrostatic charge.
• Because the enclosure of the XP2i and XP2i-DD is made of aluminium, if it is mounted in an area where the use of category 1 G apparatus
is required, it must be installed such that, even in the event of rare incidents, ignition sources due to impact and friction sparks are excluded.
CERTIFICATIONS
The XP2i has been tested and certied to comply with a variety of international standards.
This XP2i complies with the Australian Radiocommunications (Electromagnetic Compatibility) Standard 2008.
Crystal Engineering declares that the XP2i is in accordance with the ATEX Directive, the Electromagnetic Compatibility Directive, and the Pressure
Equipment Directive per our declaration(s).
This XP2i is approved for use as a portable test instrument for Marine use and complies with Det Norsjke Veritas’ Rules for Classication of Ships, High
Speed & Light Craft, and Oshore Standards.
Exia Intrinsically Safe and Non-Incendive for Hazardous Locations:
Class I, Division 1, Groups A, B, C and D, Temperature Code T4/T3C.
S
écurité intrinsèque et non incendiaire pour dangereux Lieux:
Classe I, Division 1, Groupes A, B, C et D, Code de température T4/T3C.