Operation / Technical Manual
Model 111W Hyperthermia System
Cincinnati Sub-Zero Products, LLC. 12011 Mosteller Road Cincinnati, Ohio 45241, U.S.A.
www.cszmedical.com

Table of Contents

Questions and Answers

  • M
    Michael DavisAug 19, 2025
    What to do if CSZ Medical Equipment displays ERR and sounds the alarm?
    • V
      Victoria TurnerAug 19, 2025
      If the CSZ Medical Equipment unit is operating, and the display reads ERR while the alarm sounds, this could be due to a defective water temperature sensor or a defective microprocessor board. To resolve this, replace the water temperature sensor or replace the microprocessor board.
  • C
    Cameron JohnsonAug 21, 2025
    Why does my CSZ Norm-O-Temp 111W Medical Equipment alarm with LO WATER light on when the water level is correct?
    • C
      cortezemilyAug 21, 2025
      If the CSZ Medical Equipment unit sounds an alarm and the LO WATER indicator light is on, but the water level is correct, it could be due to a disconnected 2-pin connection cable from the LOW WATER float switch or a defective low water level sensor. Try reconnecting the 2-pin connector. If that doesn't work, replace the low water level sensor.
  • J
    Jaime ColeAug 25, 2025
    How to fix a CSZ Norm-O-Temp 111W Medical Equipment unit that won't heat or pump water?
    • J
      Juan HintonAug 25, 2025
      If the CSZ Medical Equipment unit is operating, but it will not heat or pump water, the independent high limit safety thermostat may have tripped or is defective. Try recalibrating the independent high limit safety thermostat. If that doesn't work, replace the thermostat if it is defective.
  • J
    jonathanriveraAug 26, 2025
    What to do if water is not circulating in CSZ Medical Equipment?
    • F
      fernando40Aug 26, 2025
      If the CSZ Medical Equipment unit is set to operate, but the water flow indicator does not move (meaning water isn't circulating), there could be several reasons: the quick-disconnect couplings are disengaged, the water line or filter is clogged, there's a kink in the hose or blanket, the pad is clogged, the pump isn't running, or the microprocessor or power supply board is defective. Check all couplings, clear any clogs using forced air, clean the water filter, straighten any kinks, check the pump voltage and housing, and if necessary, replace the pump motor, microprocessor board, or power supply board.
  • C
    calhounnicoleAug 28, 2025
    Why do the switches on my CSZ Norm-O-Temp 111W Medical Equipment not work?
    • E
      Elizabeth HartmanAug 28, 2025
      If a switch on the membrane control panel of the CSZ Medical Equipment unit is pressed but nothing happens, the clear strip cable leading from the membrane control panel to the microprocessor control board might be disconnected, or the membrane control panel itself may be defective. Reconnect the 7-pin clear stripe cable. If the issue persists, replace the membrane control panel.
  • K
    Kevin BergAug 30, 2025
    Why does my CSZ Norm-O-Temp 111W keep turning off and on?
    • T
      Thomas LloydAug 30, 2025
      If the CSZ Medical Equipment unit is operating and it momentarily blanks before restarting, the issue might be due to low line voltage, a power interruption, or a loose connection of the power cord. Check the line voltage and ensure the power cord is securely connected.
  • T
    tylerturnerSep 2, 2025
    Why is the control panel on my CSZ Norm-O-Temp 111W blank?
    • L
      Laura MillerSep 2, 2025
      If the power switch of the CSZ Medical Equipment is pressed to the ON position, it lights up green but the control panel is blank or inoperable, the power supply board may be defective, or the cables connected to the microprocessor control board may be disconnected. Replace the power supply board or reconnect the 12-pin and 9-pin connector cables.
  • M
    Miranda HarrisonSep 4, 2025
    Why is my CSZ Medical Equipment not heating?
    • X
      xgreenSep 4, 2025
      If the CSZ Medical Equipment unit is set to operate but fluid does not heat, this could be due to water not circulating, a defective heater, or a defective power supply board. Check if the water is circulating. If not, troubleshoot circulation issues. If the water is circulating, replace the heater or replace the power supply board.
  • L
    levinejoanSep 6, 2025
    Why does my CSZ Medical Equipment display 333?
    • S
      Sara MartinezSep 6, 2025
      If the CSZ Medical Equipment unit is set to operate. When the unit goes through its test indicator cycle, the display shows a typical pattern display; e.g. all 333’s, this is due to a microprocessor control board malfunction. Replace the microprocessor control board.
  • J
    Jackie SerranoSep 8, 2025
    What to do if the silence alarm switch doesn't work on CSZ Norm-O-Temp 111W Medical Equipment?
    • J
      jcooperSep 8, 2025
      If the CSZ Medical Equipment unit is ON, and the unit sounds the alarm and the indicator lights up either LO WATER or HIGH TEMP, but the silence alarm switch does not stop the alarm, the membrane control panel has likely failed. Replace the membrane control panel.

Summary

Section 1. Introduction

General Safety Precautions

Crucial safety warnings and precautions for system operation.

System Description and Intended Use

Explains the system's purpose, components, intended use, and environment.

Physical Description and External Features

Details physical features, including front, left, rear, right, and top views.

Required Accessories

Lists essential accessories for NORM-O-TEMP® system operation.

Functional Description and Safety Controls

Explains system functions, heating, circulation, and temperature safety controls.

Section 2. Specifications and Certifications

NORM-O-TEMP, Model 111 W Features

Summarizes unit features, dimensions, and specifications.

Section 3. General Preparation of the NORM-O-TEMP System

Preparation Introduction and Unpacking

Introduces preparation steps and guides on unpacking the NORM-O-TEMP® system.

First Time Set-Up and System Test

Covers initial setup, inspection, and performing a system test.

Unit, Patient Precautions, and Bedside Care

Outlines unit, patient precautions, and essential bedside care procedures.

Alarms and Error Displays

Explains various alarms and error displays and their meanings.

Section 4. Operating the NORM-O-TEMP System

Arranging Components and System Operation

Details component arrangement and step-by-step system operation.

Concluding System Use

Guides on properly concluding system use after therapy.

Section 5. General Maintenance of the NORM-O-TEMP System

Maintenance Introduction and Reservoir Care

Introduces maintenance and details reservoir care, cleaning, and replenishment.

Water Filter, Exterior, and Blanket;Pad Maintenance

Covers water filter, exterior, and blanket/pad maintenance.

Section 6. Field Repair; Service of the NORM-O-TEMP

Service Introduction and Interior Access

Introduces service tasks and explains how to access internal components for repair.

Component Replacement Procedures

Covers replacement of key components like heater, pump, sensors, and boards.

Calibration, Testing, and Electrical Safety

Details calibration, safety tests, and electrical safety checks for the unit.

Section 7. Troubleshooting Guide

Section 8. Parts Information

Parts Ordering, Inventory, and Returns

Details ordering parts, recommended inventory, and warranty returns.

Shipping Parts and Figures

Guides on shipping parts and refers to detailed parts list figures.

Guidance and Manufacturers Declarations

Electromagnetic Emissions and Immunity Guidance

Details unit's emissions, immunity, and environmental considerations.

RF Equipment Separation Distances

Provides guidelines for separation distances from RF communication equipment.

CSZ Norm-O-Temp 111W Specifications