I-Class
Three samples are required:
Empty: Nothing in the sensor.
Gap (or Mark): The media liner, notch, or reflective mark in the sensor.
Paper: The label (and liner, if any) in the sensor.
With the correct SENSOR TYPE selected (see Section 4.2.2), perform a Standard Calibration
as follows:
Step
Action Displayed Message Comment
1 Turn ON the printer.
UNCALIBRATED
Wait briefly, about six
seconds, for the printer
to initialize.
2
Press the
Key, and
then raise the printhead
assembly.
MENU MODE
MEDIA SETTINGS
You are in MENU MODE.
3
Press the
Key.
MEDIA SETTINGS
MEDIA TYPE
You are in MEDIA
SETTINGS.
4
Press the
Key then
scroll to SENSOR
CALIBRATION.
MEDIA SETTINGS
SENSOR CALIBRATION
Press the
Key to
cancel this procedure.
5
Press the
Key.
SENSOR CALIBRATION
PERFORM CALIBRATION
You are beginning the
procedure.
6
Press the
Key, and
then press any key.
REMOVE LABEL STOCK
PRESS ANY KEY <yyy>
No media should be in
the sensor. This sets the
Empty value, where
“yyy” is the current
sensor reading.