Page 22 Tec 5 Vaporizer
August 1999 O & M Manual Part No. 1105-0100-000
7.1.2 Fluotec 5%
Effect of Flowrate at 21 ± 2° C with oxygen flowing
004A T5 OM
Flowrate (litres/minute oxygen)
Effect of Temperature at 5 litres/minute oxygen flow
005A T5 OM
Temperature ° C
4
0.2 1
2
3
5
4
% HALOTHANE
DIAL SETTING
5
3
2
1
0
5
10 15
1
0.6
0.2
5
4
3
2
1
0
% HALOTHANE
15
20
25
30 35
4
5
DIAL SETTING
3
2
1
0.6
0.2