Do you have a question about the Datex-Ohmeda Tec 7 and is the answer not in the manual?
Manufacturer | Datex-Ohmeda |
---|---|
Agent Specific | Yes |
Temperature Compensation | Automatic |
Flow Compensation | Automatic |
Agent | Sevoflurane, Isoflurane, Halothane, Enflurane |
Vaporizer Type | Variable bypass |
Flow Rate Range | 0 to 15 L/min |
Calibration | Agent-specific calibration |
Output Concentration Range | 0 to 5% |
Important safety warnings and precautions to be followed before operating the device.
Key information and recommendations regarding the vaporizer's use and documentation.
Explanation of various warning symbols and technical notations used in the manual.
Overview of the Tec 7 Vaporizer's design and intended use in inhalation anesthesia.
Description of the single control dial used for setting anesthetic agent concentration.
Explanation of the interlock mechanism ensuring safe operation with manifold systems.
Information about the label used to identify the type of vaporizer.
Step-by-step guide for securely mounting the vaporizer onto the anesthesia manifold.
Detailed instructions for attaching the vaporizer to the manifold, including lever operation and manifold preparation.
Procedure to verify that the vaporizer is properly secured and functioning correctly on the manifold.
Instructions for safely detaching the vaporizer from the anesthesia manifold.
How to adjust the control dial to set the desired anesthetic agent concentration.
Procedures for safely filling and draining the vaporizer with anesthetic agents.
Step-by-step guide for filling the vaporizer using the funnel filler system.
Instructions for draining the vaporizer using the funnel filler system.
Guide for filling the vaporizer using the Easy-Fil™ agent specific filling system.
Instructions for draining the vaporizer using the Easy-Fil™ system.
Steps for filling the vaporizer using the Quik-Fil™ agent specific filling system.
Instructions for draining the vaporizer using the Quik-Fil™ system.
Guidelines and warnings related to user maintenance and avoiding tampering with the vaporizer.
Recommended intervals for performing maintenance and safety inspections.
Detailed instructions for cleaning the vaporizer's external and internal surfaces.
Procedure for cleaning the exterior of the vaporizer with a moist cloth and mild detergent.
Steps to take if the vaporizer is contaminated with incorrect agents or substances.
Procedure for verifying the accuracy of the vaporizer's delivered agent concentration.
Information on confirming the vaporizer's calibration accuracy and factors to consider.
Methods and considerations for field checking vaporizer calibration using analytical techniques.
Information on where and how to obtain authorized repairs and maintenance services.
How the locking lever and control dial interlock for safe operation.
Overview of how the vaporizer delivers anesthetic agent vapor through gas flow.
Explanation of the bypass circuit that controls gas flow proportion.
Description of the vaporizing chamber circuit and how it enriches gas with anesthetic vapor.
Information on how the vaporizer is calibrated and the calibration certificate.
Details on vaporizer accuracy, liquid capacity, flow resistance, and operating temperature ranges.
Physical specifications including weight, depth, width, and height of the vaporizer.
Graphs illustrating the effect of flowrate and temperature on vaporizer output for various agents.
Calculations and examples for estimating anesthetic agent consumption rates.
How ambient barometric pressure affects vaporizer output and compensation.
Impact of ambient temperature variations on vaporizer output accuracy.
Explanation of how steady and fluctuating back pressures affect vaporizer output.
Effect of using carrier gases other than oxygen on vaporizer output.
Description of transient vapor concentrations after periods of inactivity.