5 Maintenance
1175-0013-000
5-5
8. Drain the vaporizer as detailed previously in Instruction 2 and measure the
amount of liquid drained off. The amount of liquid consumed should be as
follows:
Enfluratec 15.5 milliliters
Fluotec 13.5 milliliters
Isotec 15.5 milliliters
Sevotec 16.6 milliliters
Appropriate action must be taken to handle the exhaust gases and spillage.
Note
This method is designed to be a quick and easy check of vaporizer operation and,
therefore , it is somewhat imprecise. However, it is unusual for the measured liquid
consumption to vary by more than 25% of the values listed above.
Service Policy
Repairs and service procedures must be performed at a Datex-Ohmeda Authorized
Service Center. Contact your Datex-Ohmeda Service Representative or Datex-Ohmeda
Authorized Distributor for information on maintenance and shipping.