What to do if the flowmeter ball bounces in De Nora Measuring Instruments?
- AAndrew PowellAug 5, 2025
If the flowmeter ball bounces or the maximum gas feed rate cannot be achieved during normal operation of your De Nora Measuring Instruments, it could be due to several reasons. The vacuum regulator inlet filter screen might be dirty, requiring replacement of the gas inlet filter. Alternatively, the rate valve or flowmeter could be dirty and need cleaning, refer to the Service section for instructions. Another potential cause is fluctuating ejector water supply pressure leading to insufficient ejector vacuum; in this case, check and correct the water supply pressure as needed.