Dew Point Sensor Volatiles Sensor (TEV only)
Standard @ 25°C Lot # a
w
±0.003 °C a
w
±0.015 °C
13.41m LiCl 0.250
0.250
8.57m LiCl 0.500
0.500
6.0m NaCl 0.760
0.760
Steam Distilled H
2
O 1.00
1.00
Water Activity Verification Standards Within Specification Yes No
Remarks: N/A
If verification standards are out of specification, clean the instrument and follow the procedure in the
User’s Manual to perform a linear offset. Repeat the verification process with fresh standards.
Dew Point Sensor Volatiles Sensor (TEV only)
Standard @ 25°C Lot # a
w
±0.003 °C a
w
±0.015 °C
13.41m LiCl 0.250
0.250
8.57m LiCl 0.500
0.500
6.0m NaCl 0.760
0.760
Steam Distilled H
2
0 1.00
1.00
Water Activity Verification Standards Within Specification Yes No N/A
Remarks: N/A
Qualified by:
Date: Initials:
Decagon Devices, Inc June 28, 2016 11:24 AM 11

Other manuals for Decagon Devices AquaLab 4TE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Decagon Devices AquaLab 4TE and is the answer not in the manual?

Decagon Devices AquaLab 4TE Specifications

General IconGeneral
BrandDecagon Devices
ModelAquaLab 4TE
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals