DECKMA HAMBURG GmbH
Issue: 27.05.08 Instruction Manual OMD-2005 Page 7 of 25
4.0 SPECIFICATION OMD-2005
Range: 0 – 30 ppm, Trend up to 50 ppm
Accuracy According IMO MEPC. 107(49)
Linearity Up to 30 ppm better than ± 2 %
Display Green Graphic Display
Power Supply: 24 V – 240 V AC or DC
Automatic Voltage Selection
Consumption: < 15 VA
Alarm Points 1 + 2: Adjustable between 1 - 15 ppm
(Works adjustment 15 ppm)
Alarm 1 Operating Delay:
(for annunciation purpose)
Adjustable between 1 – 540 sec.
(Works adjustment 2 sec)
Alarm 2 Operating Delay:
(for control purposes)
Adjustable between 1 – 10 sec.
(Works adjustment 10 sec)
System Fault Alarm: Red LED
Alarm Contact Rating: Potential free 1 pole change over
contacts, 3 A / 240 V
Alarm Indication: Red LED's
Output Signal: 0 – 20 mA or 4 – 20 mA for 0-30 ppm
reversible, ext. Load < 150 Ω
Sample Water Pressure: 0,1 – 10 bar
Sample Flow: Approx. 0,1 - 4 l/min depend. to pressure
Ambient Temperature: + 1 to + 55° C
Sample Water Temperature: + 1 to + 65° C
Roll: Up to 45°
Size (over all): 360 mm W x 240 mm H x 100 mm D
Degree of Protection: IP 65
Weight: 7,3 kg
Pipe Connections: R ¼" Female