12.6 Removal of Iron Oxide Layer with Citric Acid
In certain cases iron oxide can be deposited inside the glass tube (brownish
surface deposit on the glass tube), depending on environmental conditions on
site. Iron Oxide can not be removed mechanically. It is recommended to remove
it chemically.
Iron Oxide can be converted into a water soluble compound using Citric Acid.
Citric Acid is commonly used to remove calcite from Coffee Machines. Fill some
Citric Acid into the drained Sample Glass Tube, add water, and let it soak
overnight (the reaction takes some time). Flush with clean water.
Additional use of tooth paste may also assist in cleaning as a last resort. Do not
use abrasive cleaners containing hard particles. Hard abrasive particles may
scratch the glass surface, permanently damaging the instrument.