Clean the oxygen inhaler and atomizing components
● Disassembly
It is located on the back of the machine; pull out the inlet cover outward
and take out the filter directly.The actual cleaning or replacement time of
filter and filter cotton shall be determined according to the actual service
time and environmental impact. If the filter is dirty and blocked, it shall be
cleaned or replaced immediately no matter how long it is used.
Note: as consumables, it is recommended to replace them regularly.
●Cleaning
clean it with light detergent, rinse it with clean water, and then install it on
the machine after it is fully dry.
They shall be cleaned once after each use. The nasal aspirator on the
snuffing oxygen inhaler shall be cleaned and disinfected after each use:
soak it in 5% potassium permanganate solution for 5min, and then clean it
with clean water or wipe it with alcohol. It is recommended to regularly
replace the oxygen inhalation tubes and atomizing components as
consumables.
Replace the fuse
After confirming that the power line plug is pulled out of the power socket,
pry the fuse cover with a slotted screwdriver (as shown in the appearance
diagram), take out the invalid glass fuse tube, install the new glass fuse
tube, and install the fuse cover back into the socket.
Prize up the fuse cover with
a small slotted screwdriver
Spare part
During use