Reset (menu Reset) restores factory settings and returns the
Gage to a known condition. The following occurs:
-
All batches, stored measurements, batch names and screen
captures are erased.
-
Calibration adjustments are returned to factory settings.
-
The last stored background measurement (
γ
1
) is erased.
-
Menu settings are returned to the following:
1. Power down the Gage and wait 5 seconds.
2. Simultaneously press and hold the (
+
) and center
buttons until the Reset symbol appears.
This returns the Gage to a known, “out-of-the-box” condition.
Set the volume of water being used for the test (3.0ml default).
Use the (
-
)(+) buttons to decrease/increase the displayed value.
The patch volume is used to calculate surface density.
Patch Volume
8
Reset
Setup Menu
Memory = OFF
Bresle Mode = ON
Auto Dim = ON
Display = None
Patch Volume = 3.0ml
Bluetooth & Stream = OFF
WiFi & Access Point = OFF
Surface Density units = mg/m
2
Temperature units = °C
Hi Res = OFF
Units
Allows the operator to change the displayed units for both
Temperature (°C or °F) and Surface Density (mg/m
2
or µg/cm
2
).
Perform a more thorough Hard Reset as follows:
Flip Display = Normal
Backlight = Normal
Auto Sync .net = ON
White on Black = ON
Sound = ON
Language = English
Battery Type = Alkaline
Bluetooth Smart = OFF
USB Drive = ON
It performs the same function as a menu Reset with the addition of:
-
Bluetooth Pairing info is cleared.
-
Menu settings are returned to the following status:
-
Date, Time and WiFi settings are not affected by either Reset.
NOTE: