DektakXTStylusProfilometerUserInstructions(MMRC)
GLAs:EmilyWarren(ewarren@caltech.edu)andHeatherAudesirk(haudesir@caltech.edu)
DocumentUpdated:4/16/2012
ThesebasicinstructionsareforgeneraluseoftheDektakXTprofilometer.Thefulluser
guidecanbefoundonthedesktopofthecontrolcomputer.
Howthesystemworks:
TheDektakXTsystemtakesmeasurementselectromechanicallybymovinga
diamondtippedstylusoverthesamplesurfaceaccordingtoauserprogrammedscan
length,speed,andstylusforce.ThestylusislinkedtoaLinearVariableDifferential
Transformer(LDVT),whichproducesandprocesseselectricalsignalsthatcorrespondto
surfacevariationsofthesample.
Oursystemisonlyconfiguredin2Dmode,whichmeans:
ManualXYtranslation:101.6x101.6mm(4x4inches)ofXYtranslation,
manualleveling
manualthetarotation
Maximumsamplethickness:50mm(1.9inches)thick
Maximumscanlength:55mm(2.16inches)
StylusTipDiameter:2microns
Whenyouarrive,theinstrumentshouldbeoff,andthecovertotheisolationchambershouldbe
closed.
Thebottomofyoursamplemustbeclean(noHgorGa/Incontaminatedsamplesallowed!)
StartUp:
1.Turnonpowersupply(Blackboxonright)
2.Press(I)buttononwhitepowercontrolbox
3.StartVision64Software.Thefollowingshouldhappen:
Thetowerassemblywillmovetoitsupperlimit.Thesystemstopswiththestylus
intheTowerUpposition.
Thescanstageinitializes.
TheVision64Welcomescreenappears,followedbytheVision64“Instrument”
tab,whichincludestheDektakXTLiveVideoDisplay
4.Afterthesysteminitializes,itissafetoloadyoursampleontothestage(Brukerrecommends
lettingthesystemwarmupfor15minutes,butwehaven’tseenthatitmakesadifferencein
measurementquality).
*Putonglovesbeforetouchinganythinginsideofthelid*
5.Checkthattheflatglassblockisfreeofdust(thisflatsurfaceiswhatallowsustomeasure
1
Question and Answer IconNeed help?

Do you have a question about the Dektak XT and is the answer not in the manual?

Dektak XT Specifications

General IconGeneral
Vertical RangeUp to 1 mm
Vertical Resolution0.1nm
Data Acquisition Rateup to 1000 Hz
Power Requirements100-240V AC, 50/60Hz
Max Sample Size200mm x 200mm
Stylus Force0.03 to 15 mg
Scan Length55mm
Stylus Tip Radius12.5 µm standard (other sizes available)

Summary

How the DektakXT System Works

System Configuration Details

Details system configuration in 2D mode, including translation limits, leveling, theta rotation, sample thickness, scan length, and stylus tip diameter.

DektakXT System Start-Up Procedure

Initial Power On and Software Launch

Steps for powering on the system, launching Vision64 software, and initial system initialization.

Sample Loading and Preparation

Instructions for loading a sample onto the stage, including positioning and centering it correctly.

Adjusting Sample Position for Scanning

Guide for manually positioning the sample using stage controls and lowering the stylus for measurement.

Calibration and Scan Parameters

Instrument Calibration

Guidance on using calibration standards to verify system functionality before measurements.

Scan Parameter Configuration

Details on setting scan type, range, profile, stylus tip, force, length, and duration for measurements.

Running and Analyzing Scans

Executing a Scan

Instructions on how to run single or multiple scans using the software interface.

Manual Leveling During Scan

Procedure for adjusting stage tilt to counteract data slope during a scan.

Data Analysis and Export

Steps for zeroing, leveling data, and exporting results using the Data Analyzer window.

Unloading and Powering Down

Sample Unloading Procedure

Steps to safely unload a sample after measurements are complete.

System Power Down

Instructions for properly shutting down the DektakXT system software and hardware.