BEFORE YOU BEGIN
Button Functions
! READ BUTTON - Reads the percent moisture
content value %MC.
@ CALIBRATION CHECK BUTTON - Checks the
meter calibration. Displays the average of up
to 100 accumulated readings; displays the
highest stored reading; erases the readings.
# SET POINT (DECREASE) BUTTON - Acts as
an arrow button when pressed after the set-
point button to decrease the set point to a
lower value.
$ SET-POINT BUTTON - Displays the current
set-point. Also acts as an arrow (scroll) button
to increase the set-point value in 1% increments.
When the battery is replaced, the meter displays its
software version for one second and then turns itself
off. After replacing the battery, you must reset the
meter as described on page 5.
CHECK CALIBRATION
hRemove the probe from the top of the meter.
hPress and hold the read button
! and check button @
simultaneously. Meter is in calibration if it displays “12”
(± .2)
If you check the calibration and the display does not read
“12,” it is likely an indication of a low battery. If this
occurs, change the battery immediately. Continued use with a
low battery may cause the meter to go out of calibration. If
you have a fresh battery and the instrument still does not
indicate an acceptable calibration, return it to DELMHORST
for service. See Service for Your Meter section.