5
TO CHECK ACCUMULATED
READINGS
This feature allows you to view the total number of all
accumulated readings for the given product you have chosen,
the average of those readings, and the highest stored reading.
hTo add a reading to the sum of all previously stored readings,
release the read button
! within 2 seconds. If you press
and hold the read button
!, the meter will repeat its read
cycle, but will not add a new reading to the storage until
the button is released.
hTo view the readings press and release the calibration
check button
@. First the meter displays the number of
accumulated readings for one second, then the average of
those readings for two seconds. Then it displays the highest
stored reading for two seconds. The total “cycle” time is
five seconds.
hTo erase all the accumulated readings, hold the calibration
check button
@ for more than 5 seconds until the meter
displays “0”.
hTo keep the accumulated readings in memory, release the
calibration check button
@ before the total cycle time is
complete.
The meter will accumulate up to 100 readings. After all
100 readings are stored, it will not add new readings until
the memory has been cleared. It will also continue to display
the average of all 100 readings as a reminder that the
memory is full.
Readings below 6% will be displayed as “0”. Those above
40% will be displayed as “99.9.” Neither will be added to
the accumulated readings or used in calculation of average or
highest reading.
TO RESET METER
hPress and release the calibration check button @.
hWithin one second, press and hold the set-point decrease
button
#.
hThe meter will reset itself and display “119.” This indicates
that the meter has been reset and that the set-point has
been set to its default value of 19%. It will also clear all
of the readings stored in memory.
F-2000 owners manualB&W:F-2000 owners manual out 4/8/09 3:18 PM
Delmhorst® F-2000 Digital Hay Moisture Tester Package, C16283 • enasco.com/farmandranch/

Other manuals for Delmhorst F-2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Delmhorst F-2000 and is the answer not in the manual?

Delmhorst F-2000 Specifications

General IconGeneral
BrandDelmhorst
ModelF-2000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals