If you check the calibration and the display does not read
“12,” it is likely an indication of a low battery. If this
occurs,
change the battery immediately. Continued use with a
low
battery may cause the meter to go out of calibration. If
you have a fresh battery and the instrument still does not
indicate an acceptable calibration, return it to DELMHORST
for
service.
CHANGE THE SET-POINT
hTo change the set-point value press the set-point increase
button
$. The meter will display the current set-point
value.
hTo scroll forward to a higher value, hold the set-point
increase button
$ while the current value is displayed
and scroll up to the set-point value desired.
hTo scroll backward through the set-point values, press and
release the set-point increase button
$. Within one sec-
ond press and hold the set-point decrease button
#.
hContinue to hold the set-point decrease button
# and the
set-point value will decrease.
hWhen scrolling in either direction, release the button to
stop at your desired set-point.
If the meter reads a %MC higher than that of the set-point,
a buzzer will sound.
TO CHECK ACCUMULATED
READINGS
This feature allows you to view the total number of all
accumulated readings, the average of those readings, and the
highest stored reading.
hTo add a reading to the sum of all previously stored readings,
release the read button
! within 2 seconds. If you press
and hold the read button
!, the meter will repeat its read
cycle, but will not add a new reading to the storage until
the button is released.
hTo view the readings press and release the calibration
check button
@. First the meter displays the number of
accumulated readings for one second, then the average of
those readings for two seconds. Then it displays the highest
stored reading for two seconds. The total “cycle” time is
five seconds.