Dentsply Sirona
Operating Instructions Orthophos E
2 Safety instructions
2.4 Condensation
67 11 001 D3352
D3352.201.12.01.0203.2019
11
2.4
Condensation
Extreme temperature fluctuations may cause condensation inside the
unit. Do not switch the unit on before it has reached normal room
temperature. See also the chapter Technical data  [→18].
2.5
Qualifications of operating personnel
The unit may only be operated by skilled or properly trained personnel.
Personnel, who are to be trained, taught, instructed or are taking part in
a general training, may operate the device only under the supervision of
an experienced person.
To operate the unit, the operating personnel must:
have read and understood the Operating Instructions
be familiar with the fundamental structure and functions of the unit
be able to recognize irregularities in the functioning of the unit and
implement the appropriate measures where necessary
2.6
Switch on the unit
The patient may not be positioned in the unit when starting it and setting
the operating mode (until the sensor has been positioned).
Malfunction may cause injury to the patient.
In case of an error that requires switching the unit off and back on
again, the patient must be removed from the unit, at the latest before
the unit is switched back on.
2.7
Radiation protection
The valid radiation protection regulations and measures must be
observed. The statutory radiation protection equipment must be used. In
order to reduce radiation exposure, Dentsply Sirona recommends using
bismuth or lead shields or aprons, especially for pediatric patients.
During an exposure, the operator should move as far away from the X-
ray tube assembly as the coiled cable of the release button permits.
With the exception of the patient, no other persons without radiation
protection are allowed to stay in the room during an exposure. In
exceptional cases, a third person may provide assistance, but not the
practice staff. Visual contact with the patient and the unit must be
maintained throughout the entire exposure.
In case of malfunctions, cancel the exposure immediately by letting go
of the exposure release button.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dentsply Sirona Orthophos E Ceph and is the answer not in the manual?

Dentsply Sirona Orthophos E Ceph Specifications

General IconGeneral
BrandDentsply Sirona
ModelOrthophos E Ceph
CategoryMedical Equipment
LanguageEnglish

Related product manuals