6 Service
6.1 Cleaning and care
Dentsply Sirona
Operating Instructions Orthophos E
92
67 11 001 D3352
D3352.201.12.01.0203.2019
6.1.3 Sterilization
WARNING
Infections can be transmitted from patient to patient.
Accessories that are not sterilized correctly can cause illness in
patients.
All accessories that are suitable for sterilization should only be
sterilized in an autoclave at 132°C (270 °F), with at least
4minutes holding time and at 2.1bar (30.5psi) overpressure.
The following accessories can be sterilized:
13 °C2 27 °F0
In addition, also use the hygienic protective sleeves; see ‘Hygienic
protective sleeves’  [→31].
WARNING
The hygienic protective sleeves are single use devices.
Contaminated hygienic protective sleeves can cause illness in
patients.
Replace the hygienic protective sleeves after each patient.