TRIAD
®
2000
Light Curing System
OPERATING/SERVICE MANUAL
For Assistance, Call DENTSPLY Trubyte
570 West College Avenue
York, PA 17405
1-800-877-0020
Technical: Ext. 4212 Clinical: Ext. 4597
Warranty/Repair Assistance: Ext. 4319 or 4581

Other manuals for DENTSPLY Triad 2000

Questions and Answers

  • N
    Nicholas WilsonAug 3, 2025
    What to do if the curing lamp does not operate on my DENTSPLY Triad 2000?
    • A
      Anthony JacksonAug 3, 2025
      If the curing lamp on your DENTSPLY Medical Equipment does not operate, several factors could be at play. First, the lamp itself might be burned out and require replacement. Second, ensure that the curing chamber access door is completely closed and that the access door interlock switch is functioning correctly. Listen for a snap when closing the door; if there's no sound, adjust the mounting bracket or replace the switch if it's faulty. Additionally, the timer module or the CURE switch might be faulty, requiring replacement. Check the continuity of the CURE switch and replace it if necessary.
  • V
    Virginia TranAug 11, 2025
    How to troubleshoot a fan that does not operate in DENTSPLY Medical Equipment?
    • C
      Candace SimmonsAug 11, 2025
      If the fan on your DENTSPLY Medical Equipment does not operate, start by checking the wall outlet to ensure the unit is receiving power; correct any issues with the outlet. Next, verify if the unit's circuit breaker has tripped and reset it if necessary. If these steps don't resolve the issue, the POWER switch might be faulty, or there could be a foreign object obstructing the fan chamber. Check the switch's continuity and replace it if faulty. Also, inspect the fan chamber for any foreign objects and remove them.
  • B
    bdouglasAug 18, 2025
    What to do if the lamp does not turn off on my DENTSPLY Medical Equipment?
    • P
      Peter Craig DVMAug 18, 2025
      If the lamp on your DENTSPLY Medical Equipment doesn't turn off, the issue might stem from a faulty TIMER switch assembly or a faulty timer module. To address the TIMER switch assembly, check its resistance across terminals #4 and #5. You should measure 100K Ohm at the full counterclockwise position and 1 megohm ±10% at the full clockwise position. If the measurements are off, replace the timer switch. If the TIMER switch assembly is not the problem, consider replacing the timer module.
  • S
    silvamarkAug 27, 2025
    Why is the turntable not rotating on my DENTSPLY Medical Equipment?
    • D
      Denise SimmonsAug 27, 2025
      If the turntable on your DENTSPLY Medical Equipment is not rotating, the problem could be a faulty TABLE switch or a faulty turntable motor. First, check the continuity of the TABLE switch and replace it if it's faulty. If the switch is functioning correctly, the turntable motor itself might be defective and need replacement.

Summary

Important Information for Triad 2000 Curing Unit

Triad 2000 Light Curing Unit General Description and Specifications

Unit Specifications

Unpacking and Installation of Triad 2000 Unit

Triad 2000 Light Curing Unit Operating Procedures

Using the Sightline for Optimal Curing

Explains how to position the appliance using the sightline for optimal curing.

Processing Procedures for Triad 2000 Unit

Step-by-step guide for placing and curing the workpiece in the unit.

Triad 2000 Light Curing Unit Maintenance

Component Removal and Replacement of Lamp

Triad 2000 Troubleshooting and Analysis

Triad 2000 Parts Listing

Triad 2000 Power Schematic

Triad 2000 Curing Unit Warranty

Appendix I: Triad II Operation

Using the Triad II Sightline

Instructions for positioning Triad II material using sightlines in the curing chamber.

Triad Applications and Material Processing Times

DENTSPLY Triad 2000 Specifications

General IconGeneral
Wavelength400-500 nm
ManufacturerDENTSPLY International
Curing Unit TypeLight Curing Unit
Intended UsePolymerization of dental materials
UsageDental laboratory
TechnologyVisible Light
Material CompatibilityCompatible with light-curable dental materials
Power Requirements120V AC, 60Hz

Related product manuals