Calibrate Both Bands
Calibration in an interference-free environment
is required after each optimization.
9.
Place the Tx in a housing on level ground 3m
from receiver as shown.
10. From the Main menu, select Calibration > 1
pt calibration and calibrate each new band.
Check Above Ground Range (AGR)
11.
Always check AGR with a tape measure to
verify depth readings on each band at
various distances up to the maximum
expected bore depth. Distance readings
should be within
±5%.
Access AGR directly on the Calibration menu.
Calibrate and check AGR for both new bands.
If you selected two bands, repeat steps 9-11 (calibration and AGR)
for the second band. An error symbol will display in the roll
indicator on the Locate screen until a 1-point calibration is
completed for the current band.
Settings Menu
Use the Settings menu to set the depth units, pitch units, time zone,
telemetry channel, roll offset, pressure units, temperature units, and
language. Set the remote display to match receiver settings.
Height-Above-Ground (HAG) Menu
HAG is the distance from the ground to the base of the receiver while it is
held. Setting HAG on the Main menu lets you take accurate below-ground
depth measurements without having to place the receiver on the ground.