Intended Use
The Panacea 6300 Bariatric Bed is intended for use within an institutional healthcare environment,
Application Environment 3 (i.e. Skilled Nursing, Transitional Care, Rehabilitation Care, Assisted Living).
Compliance with the regulations and guidelines as specified by your facility is recommended.
This bed is NOT intended for patient/resident transport. Wheels are provided only to allow movement
within the patient/resident room for cleaning or patient/resident access.
Contraindications
This bed should not be used by persons who do not have the cognitive skills to understand the
information in the owner’s manual, operate the bed or control the movements of the bed through the
control interfaces. This bed should not be used or operated by children.
General Safety
NOTE: In this manual, the terms patient and resident refer to a resident of a Nursing Home, Long Term
Care, Assisted Living or Rehabilitation facility.
Danger
NEVER allow anyone under the bed at any time.
DO NOT let any body parts protrude over the side or between parts, especially when the bed is
being operated.
Danger Shock Hazard
DO NOT plug anything into the control box of bed (i.e. pendants and actuators) while mains cable
(power cord) is plugged into the wall outlet.
Any cords or tubing used on or with this bed MUST be routed and secured properly to ensure they do
NOT become entangled, kinked or severed during normal operation of the bed.
DO NOT roll the bed over any power or pendant cords.
This bed is equipped with a three-prong grounding plug for protection against possible shock hazard.
DO NOT, under any circumstances, cut or remove the grounding prong.
DO NOT open any actuators, control boxes or pendants. Service is only to be performed by authorized
service personnel. If unauthorized service is performed on any components, the warranty is void.
Inappropriate handling of power supply cords can create a shock hazard.
When routing cables from other equipment in the medical bed, precautions shall be taken to avoid
squeezing those between parts of the medical bed.
DO NOT use an extension cord.
DO NOT allow the cord, electrical outlets, and electrical control box or hand pendant to become
wet or submerged.
DO NOT operate the bed if any electrical component, such as the power cord, electrical outlet,
connections, motor/actuator or mechanical component, has malfunctioned or has been damaged
in any way.
To avoid risk of electric shock, this equipment must only be connected to a supply mains with
protective earth.