18
▸
Do not apply Icon-Infiltrant under direct oper-
ating light as this may cause the material to set
prematurely.
8.
Screw anew Proximal-Tip onto the Icon-Infiltrant
syringe, and introduce the application tip into the
interdentium. Be sure to align the green side of the
Proximal-Tip with the area to be treated.
▸
The material can only be dispensed on the green
side of the Proximal-Tip.
9. Apply an ample amount of Icon-Infiltrant onto the
lesion site (1½ to 2 turns of the shaft).
10.
Let Icon-Infiltrant set for 3 minutes. The Infiltrant
will be activated by slightly moving the applicator.
Add material if necessary.
11. Remove the application tip from the interdentium.
Remove excess material with dental floss.
6. Remove the application tip from the interdentium.
Suction off Icon-Etch and rinse with water for amin-
imum of 30 s. Dry carefully with oil-free and water-
free air.
Drying
For best treatment results it is necessary to dry the
lesion in-depth.
7.
Screw the application tip onto the Icon-Dry syringe,
apply an ample amount of material onto the lesion
and allow to set for 30 s. Dry thoroughly with oil-
free and water-free air.
Infiltration
The lesion can be penetrated in-depth by the Infiltrant
after complete drying.