20
6.
Screw the application tip onto the Icon-Dry syringe,
apply an ample amount of material onto the lesion,
and allow to set for 30 s.
Drying
For best treatment results it is necessary to dry the lesion
in-depth.
7.
After the visual check dry the lesion thoroughly with
oil free and water-free air.
Infiltration
The lesion can be penetrated in-depth by the Infiltrant
after complete drying.
8.
Screw anew Smooth Surface-Tip onto the Icon-Infil-
trant syringe.
Do not apply Icon-Infiltrant under direct oper-
ating light as this may cause the material to set
prematurely.
Note: Unless white spots are treated early, i. e. 1 to 2
months after bracket removal, it is recommended to
repeat the etching step up to three times.
5.
Suction off Icon-Etch and rinse with water for at
least 30 s. Dry with oil-free and water-free air.
Visual Check
For best treatment results it is necessary to dry the
lesion in-depth.
In this step apreview of the final result is shown.
When wetted with Icon-Dry, the whitish-opaque
coloration on the etched enamel should diminish.
If this is not the case repeat the etching step once
or twice for 2 minutes each, and rinse and dry the
teeth again (steps 3-5).