USER'S MANUAL UNIVERSAL CENTRIFUGE CENTRIC 150
Page 4
2.4 SECURITY FUNCTIONS
- Solid casing and centrifuging chamber
- Safety cover mechanism
- Imbalance detection
- Self test procedures
Design was made considered the following standards:
- IEC 1010-2-20
- BS 4402
- UV VBG7Z
- DIN 58970 parts 1,2 and 4
3.0 INSTALLATION
Installation must be made by following steps:
3.1 CENTRIFUGE PLACING
Apparatus must be placed on horizontal smooth base, without swinging.
Centrifuge is constructed for indoor use only ! It should be placed at least 300 mm from the
wall, other devices and accessories to prevent injury and damage. Do not place it near heating
elements !
3.2 ATTACHING THE POWER CORD
Fit the mains cord, which is included in the delivery in-to mains receptacle. Connect the
other end of the cord to a grounded wall-socket, protected with 16A fuse. (look at chapter 2.3)
3.3 PRIMARY EQUIPMENT
Following equipment is enclosed to centrifuge:
- User's Manual
- Warranty
- Special wrench for removing rotor
- Special tool for rotor removing
- Special wrench for manual opening cover
- Spare fuses 5 A - T or 6.3 A – T for 115V version
3.4 COVER OPENING
Cover can be opened only when device is powered by mains power and rotor stops.
You can open the cover by turning button on right side in clockwise direction. If cover is opened,
light indicator COVER blinks.