USER'S MANUAL UNIVERSAL CENTRIFUGE CENTRIC 150
Page 8
Rotor RA 24 M
Code 455.270
Rotor RH 8 M
Code 455.280
Adapter code (set 12 ) 455.285 455.287
Volume ( ml ) 0,25 1,5 0,25 - 0,4 1,5 - 2
N
o
positions 24 24 24 16
Dimensions of test tube
D X L ( mm )
6 x 30 11 x 40 6 x 30
6 x 50
11 x 40
11 x 50
Max. radius ( mm ) 65 80 60 / 80 70 / 80
Max. speed ( min.
-1
) 13000 13000 13000 13000
Max. centr. accel ( g ) 12280 15115 15115 15115
Max. kinetic energy 4020 Nm
table 3
Rotor HEMATOCRIT measure plate
Code 455.290
Volume ( ml ) 24 capillary 1,2 x 75mm
Max. radius ( mm ) 92
Max. speed ( min.
-1
) 13000
Max. centr. accel. ( g ) 17380
Max. kinetic energy 3245 Nm
table 4
5.0 CENTRIFUGING
5.1 CENTRIFUGE CONTROL
Microprocessor controller board supervise all functions in device. Regulate motor speed, read
keyboard, display all necessary parameters and control all security functions. Operator manage
device with four keys keyboard, LC display and three light indicators (figure 5)
figure 5
Operator can set following parameters:
- working time
- rotation speed
- acceleration / deceleration level
All parameters are stored in memory unit (EEPROM).