Proper protection, such as gloves, eye wear, and aprons, are recommended when disassembling a Dosatron.
Step 3: Slide the ferrule back.
Step 2: If there is still no suction, unscrew the suction hose nut
and remove the suction hose.
NOTE: To troubleshoot your Dosatron D14MZ3000, you will need
a pair of needle nose pliers and a strap wrench.
Step 1: Take a look at the strainer. If it’s clogged, just clean it and
re-test your Dosatron unit.
D14MZ3000 Troubleshooting - Part 1
This D14MZ3000 - 14 GPM Troubleshooting Part 1 will cover: The injector is clicking,
but not drawing the concentrate.
Scan the QR code to
watch our D14MZ3000 -
Troubleshooting video.