USERMENU
TTSeriesLR Manual
Page05
TheUserMenuisusedtoinputuservaluesandchangedefaultsettings.
ToaccesstheUserMenu,switchontheinstrumentwhileholdingdownENTER.
Page05
SetClock: Tosettheclocktolocaltimeandinputthedate inthechosenformat.
AutoZeroTime: Thisisfactorysetat60se
conds.
(forwhichtheinstrumentspecificationsarebased)
Changingthedefaultsettingwilloverridethis.
Inputting0secondswillswitchofftheautozerofunction.
AmanualzeroingfacilityisavailableseeKeypadControlspage02.
Page06
SwitchOffPeriod: Tosavebatterytheinstrumentwillsw
itchoffafter10minutes providingthatno
keysarepressed.Thiscanbeoverriddenbychangingthedefaultsetting.
RS232Period: Toprintoutrawdataasmeasurementsarebeingmade,eithertoathermal
printerortoaPCviatheHyperterminal.
Note: DPMeasurementnolongersupplyth
ermalprinters.
BackLightPeriod: Toalterthebacklighttime. Usey andz toswitchonthebacklight.
Temp/Pressure: Tochangetheairstreamtemperatureandbarometricpressure.
Airstreamtemperaturesetat16°CandBarometricpressuresetat1000mbar.

Page07
SelectDateFormat: SetthedateformattoEuropeanorUS
A,thetemperatureunitsto°Cor°Fand
thebarometricpressureunitstomborinHg.

Selectingfromalist:
Enteringavalue:
Usey andz toscroll.PressENTERtoselect.
UseyandztoseteachdigitorvaluethenpressENTER.
SetCl
ock SelectDateandTime
AutoZeroTime Press ENTER Minutes? 16
SwitchOffPeriod Hours? 14
RS232ResultPeriod Date? 26
BackLightPeriod Month? 10
Temp/Pressure Year? 10
SelectDateFormat UserMenu
Exit
SetClock AutoZeroPeriod
AutoZeroTime 060Seconds
SwitchOffPeriod PressENTER MaxSetting: 600
RS232ResultPeriod MinSetting: 010
BackLightPeriod Toswitchoffinput: 000
Temp/Pressure UserMenu
SelectDateFormat
Exit